| CyBi®-Lumax flash HS | CyBi®-Lumax flash HT |
| Ultimate sensitivity ...achieved by the use of a proprietary lens with a high optical aperture and excellent photon collection efficiency (CyBi®-Luminox lens), and the beneficial reproduction scale. | Ultra high throughput ...achieved by the use of a 16-channel dispensing head and a detection technology that measures light emitted from all wells of a plate at the same time. |
| Detection modes | Flash luminescence Glow luminescence | Flash luminescence Glow luminescence |
| Detector | Image intensifer plus CCD camera, 2-D single photon counting CyBi®-Luminox lens | Image intensifer plus CCD camera, 2-D single photon counting Leica® lens |
| Detection Limit | 60 fg / well Aequorin 200 fg / well Luciferase | 0.3 pg / well Aequorin 1.0 pg / well Luciferase |
| Dispenser | Two 8-channel non-contact dispensers, optimized for cells, with cell recirculation | Two 16-channel non-contact dispensers, optimized for cells, with cell recirculation; optional 8-channel heads for 96-well plates |
| Volume Range | 0.5 µL – 50 µL | 2 µL – 25 µL |
| Plate Formats | 384- and 1536-well | 96-, 384- and 1536-well |
| Reading Time | 384: 5.5 minutes 1536: 11.0 minutes | 384: 1.0 minutes 1536: 3.0 minutes |